Sanitary Tuning Fork Density Meter For monitoring density and temperature -50℃~+200℃
working principle
D= uncalibrated density of the measured medium (kg/m³)
T- vibration frequency (μs).
K0, K1, K2= constants
In the process of density detection, the NYDE-CR plug-in density meter can automatically compensate the effect of temperature on the density (D) of the measured medium, while the effect of pressure on the density is not significant.
design feature
• Suitable for density measurement of storage tanks, pipelines, flowing or static media. Suitable for measuring high temperature, low temperature and high viscosity media containing impurities.
• A microprocessor-based electronic switching device that integrates signal processing, calculation and diagnostic functions.
• Continuous on-line measurement of liquid density and temperature, with high precision, more safety and reliability, and can be directly used for production process control.
• Can communicate with PC directly through RS485 interface, users can directly configure nodes online, diagnose faults, and record data.
• While measuring density and temperature, basic density, concentration percentage, mass percentage and volume percentage can be calculated.
•Excellent overvoltage protection structure, excellent temperature stability, fully welded seal, solid and reliable design.
•Easy to install and use, pull in the liquid to display the reading.
• Simplified maintenance, no need for regular cleaning.
• Calibration does not require a standard reference source, laboratory calibration and process interruption.
• The intrinsically safe type can be used in hazardous locations, and the hygienic type can be installed in food production sites.
technical parameter
typical application
|
Tank Tube Density Measurement
|
Product number
|
Enjoy - CR |
shell material |
cast aluminum |
Measuring range
|
0.5~2.5g/cc(500-2500kg/m³) |
Calibration range |
0.5~2.5g/cc(500-2500kg/m³) |
accuracy |
0.001 g/cm3 |
Signal output
|
4~20mA,0~1000HZ,RS485,Modbus RTU |
Operating temperature
|
-50℃~+200℃ |
greatest pressure |
207bar(3000psi) |
Contact material |
316L, Stainless Steel, Hastelloy
|
forked body |
Standard, PTFE, Electropolished
|
power supply |
24VDC≥50mA |
connect |
ANSI150~1500RF |
DN50PN16/DN50PN40 |
BathroomIDF/RJT |
Internal temperature sensor |
digital sensor |
Protection class
|
IP65, IP67, IP68 (optional)
|
Explosion-proof grade
|
ExdⅡCT6 ,ExmbⅡCT6
|